HIV GENOSURE

GenoSure Prime (HIV-1 Genotype)

EPIC Test Procedure Code: LAB4021

Performing Lab:
Referral Laboratory
Container Type:
Lavender top (EDTA) tubes or plasma preparation tubes (PPT™)
Specimen Type:

Plasma

Preferred Volume:
5 mL
Collection Procedure:

Collect blood in three tubes, 5 mL each of lavender top (EDTA) or PPT™ tubes.

Specimen Processing:

CENTRIFUGE THE TUBES WITHIN 6 HOURS OF COLLECTION, REMOVE THE PLASMA AND TRANSFER TO A SEPARATE SCREW CAP TUBE AND FREEZE.  IF THE SAMPLE IS COLLECTED IN PPT™ TUBES, THERE IS NO NEED TO REMOVE THE PLASMA AND PLACE IN A SEPARATE TRANSPORT TUBE. IT DOES NEED TO BE SPUN WITHIN 6 HOURS OF COLLECTION.

Store and Transport:
Frozen
Stability:

Frozen - 2 years

Unacceptable Condition:

Insufficient quantity, green top (heparin) tube used, unspun  PPT™ tube, or unfrozen samples.

Limitations:
This procedure may not be successful when the HIV viral load is <500 copies/mL plasma.
CPT Codes:

87900 - HIV Genosure Prime 3 (EAP 30035079)
87901 - HIV Genosure Prime 2 (EAP 30035077)
87906 - HIV Genosure Prime Interp (EAP 30035078)

Method:
Polymerase chain reaction (PCR) amplification and DNA sequencing
Test Includes:
Includes the HIV-1 Integrase
Notes:

Tests four drug classes (NRTIs, NNRTIs, PIs, and INIs).

Lab Personnel

Please login to see billing, ordering, and reporting instructions.