ANTI XA UFH
Heparin Anti-Factor Xa Unfractionated Assay, Plasma
EPIC Test Procedure Code: LAB2185
Platelet Poor Plasma
1. Centrifuge specimen in a centrifuge that has been validated to ensure platelet poor plasma. Centrifugation should take place within 1 hr of collection.
2. If plasma is to be frozen:
a. Immediately transfer the plasma to a stoppered, plastic tube and centrifuge again for 10 minutes.
b. Transfer the supernatant plasma into a screw-capped, plastic vial and send frozen.
3. If unable to centrifuge, forward unprocessed sodium citrate whole blood (minimum volume: 2 mL) to the laboratory promptly at room temperature only.
The unspun specimen must arrive within 30 minutes of collection. Anti-Xas will not be run on specimens > 2 hours old.
85520 - Anti Factor XA Assay UFH (EAP 30021175)
UFH
- Therapeutic range
0.3 - 0.7
IU/mL
(or as specified by order set or physician).
UFH
- Critical values (automatic call-back):
> 1.1
IU/mL
Used to monitor unfractionated heparin (UFH).